Flame Atomic Absorption / Emission Spectrometry
Analytes currently available include: sodium, potassium, calcium, magnesium and lithium. Future plans exist to transfer these methods onto an Inductively Coupled Plasma Mass Spectrometer (ICP-MS) platform.
| Analyte | Standard (certfied purity) | Control Material |
|---|---|---|
| Sodium | Na CL, Sigma >99.999% | SRM 909al, 909all, 909b1, 909bll |
| Potassium | K Cl, Sigma >99.999% | SRM 909al, 909all, 909b1, 909bll |
| Calcium | CaCO, Sigma >99.995% | SRM 909al, 909all, 909b1, 909bll |
| Magnesium | Mg Gluconate, Sigma >98% | SRM 909al, 909all, 909b1, 909bll |
| Lithium | LiCO2, Sigma >99.997% | SRM 909al, 909all, 909b1, 909bll |
| Cation | Sample pre-treatment | Technique | Wavelength (nm) |
|---|---|---|---|
|
Sodium |
1/2025 dilution Ionisation suppressant (CsCl)* Surfactant (0.1% Triton) |
Emission | 589.6 |
| Potassium |
1/450 dilution |
Emission | 766.5 |
| Calcium |
1/25 dilution |
Absorption | 422.7 |
| Magnesium |
1/150 dilution |
Absorption | 285.2 |
| Lithium |
1/15 dilution |
Absorption | 670.8 |
*2000 PPM, **1000 PPM
Sample Handling
All samples are diluted using an automated robotic workstation directly from the primary sample.
| Analyte | Proposed AEmax(%) | Actual AEmax(%) | Proposed Biasmax(%) | Actual Biasmax(%) |
|---|---|---|---|---|
|
Sodium |
1.3 | 0.8 | 0.5 | 0.2 |
| Potassium |
2.0 |
1.7 | 0.8 | 0.7 |
| Calcium |
1.7 |
1.1 | 0.5 | 0.4 |
| Magnesium |
1.7 |
1.5 | 0.5 | 0.5 |
| Lithium |
2.4 |
1.9 | 1.2 | 1.9 |